Jet-X at MSSL |
KFRD |
KFRD consists of two instruments FRD-A (carriage A) and FRD-B (carriage B). Instruments FRD-A and -B are fully identical. They are two independent position sensitive x-ray spectrometers. Within the detectors, the electrode planes are in two layers. The first layer is position sensitive and spectrometric, the second one - only spectrometric. The second layer is also used for anticoincidence system with the aim to reduce the charge particle background in the MWPC. All electrode planes are mounted on ceramic frames.
Instrument performance The table lists the instrument performance.
Detector | MWPC filled with 85%Xe+10%Ar+5%CO2 at 1.5 atm |
Energy Band | 2 - 25 keV |
Entrance Window | 50mm Be depth 120mm |
Efficiency/effective area | 0.93/1100 cm2 at
5.9keV 0.44/35 cm2 at 22.1keV |
Energy resolution | 17% at 5.9keV, 9% at 22.1 keV |
Spatial resolution | 0.5mm at 5.9keV (10 arcsec in focal plane) |
Angular resolution | 2 arcmin (limited by mirrors |
Timing | 1/4096 - 250 microsec |
Dead Time | 50 microsec |