XMM RPS
Users' Manual
Next: Example observation of a
Up: Observation examples
Previous: Observation examples
Example for an observation of a faint extended source
As an example for a faint extended X-ray source one might consider
observing a cluster of galaxies, at relatively low redshift, z.
Input of standard information (target name, catalogued position
etc.) in XRPS is trivial. Other input parameters require some
more thought:
- Choice of prime instrument
- There might be a cooling flow at the centre. If moderate resolution
spectroscopy is most important for the proposed science, the observer
might want it to be imaged in the 'aim point' of the EPIC pn camera.
Alternatively, if high-resolution spectroscopy is intended, (s)he may
want to centre it on the RGS-1 instrument (which is the RGS unit
with the highest energy resolution).
- Science mode of the prime instrument
- If the source is weak, all EPIC cameras can be assumed to have no
problems with photon pile-up. In that case, they can all be used in
the standard FULL WINDOW imaging mode. RGS could be used in its
standard SPECTROSCOPY mode.
- Total integration time requirement
- Based on PIMMS observers can convert ROSAT, ASCA or other known
flux and band data (if known) into the XMM band (0.1-15 keV), and
enter these, together with, e.g., X-ray spectral model = Raymond-Smith,
kT = 6 keV, N(H) = 3e20 cm-2 and the lower and upper limit of
the energy band over which the X-ray flux was observed into XRPS.
- Length of observation vs. visibility constraints
- Users must check that the requested observation fits into a continuous
visibility period of the XMM orbit, using the XMM Target Visibility Tool. In case the
required total integration time is longer than the longest possible
visibility window, the observation must be split into an adequate
number of individual observations.
- Pointing coordinates
- Assuming that the cooling flow is located at the core of the cluster,
no boresight coordinates need be entered and thus the target coordinates
will automatically be propagated into the boresight fields. Otherwise,
the boresight would be chosen to be directed towards the location of
brightest/most important feature to be observed. The best data quality
will be achieved in the aim point of the prime instrument.
- Avoidance of nearby bright sources
- Optical and X-ray catalogues should be searched for nearby bright
sources which might lead to contamination of either the X-ray (e.g.,
RGS spectral overlaps) and/or optical/UV observations. If any such
sources exist and must be avoided, this might lead to a position
angle constraint (which in turn makes the observation ``fixed'' in
time).
- Science modes of the other instruments
- The expected RGS count rates are lower than those for EPIC. There will
thus be no need for fast readouts and RGS would be used in its standard
SPECTROSCOPY mode.
Assuming that the cluster fits into the OM's 17' FOV, a default
configuration (matching the choice of either EPIC pn or RGS-1 as
prime instrument!) would be chosen (``EPIC pn IMG'' or ``RGS 1 IMG'').
- EPIC filters
- Since the source is assumed to be weak (both in X-rays and the optical/UV)
the ``THIN'' optical blocking filter can be used. Observations without
any filter might be contaminated by optical radiation (diffuse background
or bright optical sources) compromising the accuracy of the EPIC energy
calibration.
- RGS readout sequence
- In case of standard spectroscopy observations there is no need to change
anything in the RGS readout sequence. CCDs 1-9 will then be read out
sequentially.
- OM brightness limit
- Before planning details of OM observations, users should check for the
presence of bright optical/UV sources within the OM's FOV. There should
be no source in the FOV that violates the brightness constraints tabulated
in UHB Table 19
. If any such source should exist (and it cannot be
avoided by, e.g., reducing the OM FOV by using the magnifier as optical
element), the OM MUST be put in the ``GO-OFF'' mode (which corresponds
to the blocked filter position) to protect it from radiation damage.
- OM filters
- If the observer has no particular interest in specific wavelength
ranges, the recommended OM filters (as listed in the UHB section
on OM default configurations
) should be chosen.
- Length of exposures
- All X-ray observations of faint sources can be obtained in a single
exposure covering the entire duration of the observation.
The OM exposure times should be chosen according to the explanations
in § 5.3.3.5 and the OM chapter
of the UHB. An example
for how OM default configurations
work is shown above, in § 3.3.
Next: Example observation of a
Up: Observation examples
Previous: Observation examples
European Space Agency - XMM Science Operations Centre