XMM RPS
Users' Manual
Up: Observation examples
Previous: Imaging observation of a
Timing observation of a variable source
Consider observing a bright point source, as above (like, e.g., an unresolved
AGN, binary or stellar object), but now with a special interest in high
time resolution measurements. In addition to trivial input like, e.g.,
target name, catalogued position, etc., input parameters to be chosen
are:
- Choice of prime instrument
- The prime instrument is chosen according to the importance of data from
either type of XMM instrument: either EPIC pn, if imaging with moderate
resolution spectroscopy is crucial, or RGS-1, if the highest possible
spectral resolution must be achieved, or even OM, if grism observations
with simultaneous flux monitoring via a fast mode window covering the
zero-order image are mandatory. Let us, for the time being, assume that
EPIC pn is prime.
- Science mode of the prime instrument
- For high time-resolution observations the EPIC pn TIMING mode is a
suitable choice.
- Total integration time requirement
- Based on PIMMS observers can convert ROSAT, ASCA or other known
flux and band data (if known), and enter these, together with, e.g.,
X-ray spectral model = power law,
= 0.7, N(H) = 3e21
cm-2 and the lower and upper limit of the energy band over which
the X-ray flux was observed.
- Length of observation vs. visibility constraints
- Users must check that the requested observation fits into a continuous
visibility period of the XMM orbit, using the XMM Target Visibility Tool. In case the
required total integration time is longer than the longest possible
visibility window, the observation must be split into an adequate
number of individual observations.
- Pointing coordinates
- No boresight coordinates need be entered to centre the target on the
prime instrument, because the target coordinates will be propagated
automatically into the boresight fields, if no other values are
provided. The best data quality will be achieved in the aim point of
the prime instrument.
- Science modes of the other instruments
- Based on the nature of the example, one can assume that all instruments
will be operated in their fast modes. EPIC pn reaches a time resolution
of 0.03 ms in its TIMING mode, MOS reaches a resolution of 1.5 ms. Note
that the pn camera in its Small Partial Window mode already reaches a time
resolution of 6 ms, which would at the same time render possible imaging
of the target. RGS would be operated in the HIGH TIME RESOLUTION mode
and OM in the fast mode default configuration EPIC pn IMG FST.
- Avoidance of nearby bright sources
- Optical and X-ray catalogues should be searched for nearby bright
sources which might lead to contamination of either the X-ray (e.g.,
RGS spectral overlaps) and/or optical/UV observations. In particular
when the EPIC TIMING mode and also the RGS HIGH TIME RESOLUTION
mode is used, one must ensure that nearby sources do not contaminate
the target data. There must be no nearby source in the same column
of the EPIC cameras as the science target. However, note that the
two MOS cameras are mounted orthogonal to each other. For the RGS
there must be no source along the dispersion direction of the target
spectrum. If any such sources should exist and must be avoided, this
might lead to a position angle constraint (which in turn makes the
observation ``fixed'' in time).
- EPIC filters
- Using the instructions provided in the UHB on EPIC filters
, the user
must decide which optical blocking filter suppresses optical loading in
the soft part of the X-ray passband sufficiently and at the same time
has minimal impact on the proposed science. Observations without any
filter, especially of bright sources, should not be conducted, because
they will be contaminated by optical radiation compromising the accuracy
of the EPIC energy calibration.
- RGS readout sequence
- The user must determine whether one single CCD (if yes, which one) or all
nine RGS CCDs shall be read out. The whole spectral range (all 9 CCDs)
can be read out in about 150 ms.
- OM brightness limit
- Before planning details of OM observations, users should check for the
presence of bright optical/UV sources within the OM's FOV. There should
be no source in the FOV that violates the brightness constraints tabulated
in UHB Table 19
. If any such source should exist (and it cannot be
avoided by, e.g., reducing the OM FOV by using the magnifier as optical
element), the OM MUST be put in the ``GO-OFF'' mode (which corresponds
to the blocked filter position) to protect it from radiation damage.
- OM filters
- If the observer has no particular interest in specific wavelength
ranges, the recommended OM filters (as listed in the UHB section
on OM default configurations
) should be chosen.
- Length of exposures
- Targets with extremely high count rates might lead to very high telemetry
rates. The limits listed in the UHB section on EPIC modes
apply. If
these are not exceeded, X-ray observations can be obtained in a single
exposure covering the entire duration of the observation.
The OM exposure times should be chosen according to the explanations
in § 5.3.3.5 and the OM chapter
of the UHB. An example
for how OM default configurations
work is shown above, in § 3.3.
Up: Observation examples
Previous: Imaging observation of a
European Space Agency - XMM Science Operations Centre