Wednesday 25 August 2004 18:30 Unpack at ISAS ************************ Thursday 26 Aug 2004 Set up EGSE Remove structure from transport container Connect EGSE, ICU and structure str4 measured at 1099 str7 measured at 1096 15:15 Check connections msslfp on FSRUN0183 15:23 run full functional test FFT_26_08_no1.doc 16:00 During short functional test mhc nacked during mhc_init14 and icu shows ground abort for the sequence. set_delay 1000.0 resume short functional test mission data header incorrect, most parameters zero. Test stopped. Shut down egse pc. Reseat mdp cable. Boot egse pc Shut down egse pc - smell of burning mdp cable burnt 16:40 mdpsim cable removed mdp cable not plugged in, egse pc booted FSRUN0184 egse pc shut down, pc ok connect mdp cable boot egse pc FSRUN0185 psu on 28V, 2.98A output on 0.03 tci-b on 0.29A, housekeeping received select eeprom 0 standby 0.70A interface ok icu off mdp halt egse off 19:30 egse pc cleaned inside with ipa egse pc turned on, mdp cable to icu unconnected Takes long time to boot. Turn off. Inspect. Take cover off, move pci 7200 to original slot. Turn on. Nothing. Take cover off. Power on. OK. Power cable to motherboard loose. Pushed in. All ok. MDP cable in. Boot ok. egse on FSRUN0186 no activity 19:40 egse off ********************************************************* Sunday 29 Aug 2004 mdpsim card and pci nudaq 7200 cards replaced with other ones. New mdpsim card cable inserted. Pc boots ok. Shutdown and transfer to cleanroom. Power up run FFT_29_08_04_no1 SFT finished 22:36 ROE finished 23:28 MHC finished 00:37 step 6, mhc didnt respond, mhc rebooted itself or was rebooted by the icu, 1 nack, ec = 2 after mhc came back run mhc_init14 re-run step 6 ack 33, nack 1 looks like mhc reboots, command header is e881 Then mhc "goes away" again, sequence is aborted by the script when nothing comes back from the mhc When mhc comes back, get 5 cmds read, 3 acks, 1 nack (ec = 2), cmd header 0xa09c ack,nack 3,1 run 'fm_integration/fm_mhc_qcm6s_test' 6,1 58 0x1000 59 0x005c 60 0xf9e3 8,1 58 0x2000 59 0x006a 60 0x5a48 6 seconds works ok. something wrong with 40 second test(with prom code??) step 9 completed ok step 10 gives shutter abort for sequence fm_light_test run mhc_init14 re-run step 10 02:11 Finish FFT run fm_mhc_qcm6s_test 58 0x1000 59 0x005d 60 0x0e9d 58 0x2000 59 0x006a 60 0x5e87 run 'fm_integration/fm_mhc_qcm40s_test' nack 1 error code = 2 58 0x1000 59 0x0262 60 0x2ba0 58 0x2000 59 0x02b9 60 0x2845 dump_prom 0xc0000,0x4000 mem_data.0017 02:23 standby icu off egse off ********************************************* 6 September 2004 Sequence 1 Sequence 2 15 October 2004 Sequence 3 Sequence 4 Friday 12 November 2004 Functional Test Wednesday 17 November 2004 XRT Flare trigger testing Thuraday 18 November 2004 Performance Test 1 Friday 19 November 2004 Performance Test 2 ******************************************** ISAS, 19 April 2005 HH, JAT, MCRW Am. Items cleaned and brought into cleanroom Egse set up and booted ok. 14:25 Purge unplugged from structure shipping container Purge flow rate gauge at maximum Structure removed from container and placed on two tables in cleanroom 14:49 Purge connected to structure 16:00 Savers put on structure connector panel 16:07 Opened single bag of icu 16:10 Savers put onto icu Tape glue residue noticed on structure handling frame (mirror end)(JAT) 17:30 Some cables attached to icu. Process stopped when reaslised there is no mdp simulator card and cable in the mdpsim computer. Probably at MSSL. ****************************** Friday 22 April 2005 JAT, MCRW 09:30 Egse mdpsim card inserted and connected. pc turned on ok. pc off. Structure connected. 10:25 Start full functional test Egse on No packets received after power on. Power off, check cables Power on, still no packets Power off. Test aborted, egse pc off. 13:05 mdp cable removed from icu 13:12 Breakout box attached to mdp cable egse on fsrun0189 No packets from icu Looks like request packets are being transmitted bt the mdpsim card ok 13:20 egse computer off unplug mdp cable from pc and buzz through Reseat the mdpsim card in egse pc Mdp cable buzzed through ok except 13 and 18 Bob now at end of mdpsim end of mdp cable. Looking at status interface at signals coming from the icu 13:45 egse on fsrun0190 28.01v 2.98 current limit icu on no packets oscilloscope showing packets being sent by the icu all 6 signals are present 13:51 egse off **************************** 23 May 2005 JAT, HH, MCRW 09:30 Set up egse computers in floor 7 cleanish room. Insert set A mdpsim/nudaq 7200/cable. msslfp does not power up. Remove set A cards. Msslfp powers up. Turn off msslfp and re-insert set A cards. Msslfp boots ok. Loopback on mdp socket. Run cmdst_test with no errors. Egse computers powered down and transported to integration building main cleanroom with icu etc. Structure removed from container. Purge line disconnected for about 3 minutes. Egse set up. 16:00 Cables attached between structure, icu and egse 16:42 msslqa off, power disconnected from wall socket. **************************************** Tuesday 24 May 2005 JAT, MCRW 10:00 Screws holding back plate to mdpdim card removed. Egse booted. FSRUN0191 10:11 bob connected Perform EIS_fm_software_upgrade_sp-053-01_24_05_05 (Perform EIS_FFT_24_05_05) Bad image data, CAM_IF_ERROR 0x08 -> 0x1018 during readout, HSL stat 0x78FF in md header. 11:04 Manual, standby, icu off Nothing obvious at connector panel. Pin 3 of HSL cable disconnected at cnp end - seemed to have been yanked... 12Mohm measured between icu common and structure common. 11:22 msslfp powered down Clam shell pressure is 1.77 torr 12:41 pin 3 cnp2 to gnd ok (1.3ohms meas) (0 second time) 1k094 between icu common and structure common (may have been disconnected from icu last time measured before lunch) Disconnected hsl at icu end. measurement is now high impedance 12:55 Broken wire found on the "outside" of the connector feed through - pin 9 (roe_sci_dat-) 13:06 Cables removed from connector panel. Cables removed from ICU. TV harness and external adaptor put away. Check EMC harnesses to ICD and 'Hiro FMharness_-3-9-3.ppt' ICU3 MDP cable OK. Note pin 13 overall screen is not fitted. Pin 1, also overall screen, is fitted. Agrees with FM harness. ICU1 WIR08 Power cable Note not connected are pins 16, 18, 15, 17, 1, 14, 13, 25, 21, 22. The harness agrees with the FM harness. ICU7 HKU tempsensors, OK CNP6 HCE Survival ccts, Note not connected are pins 1, 13. Agree with FM harness CNP5 HKU STR temp sensors. Note connected is pin 1, Agrees with FM harness. ICU6 CNP4 Bake out heaters. Note not connected are 1 and 8. Agress with FM harness ICU2 CNP1 RS422 links, As FM harness accept 14 not connected to ICU end but loops back to pin 11. This as per the new grounding arrangement of the MHC ICU5 CNP3, secondary CAM power and primary for MHC. Screen pins 1 and 19 at ICU end do not go to 1 and 19 but 20 and 37 (wrong row). This will have been wrong for the EMC test. Screen will have been grounded at one end but not at the other. FM harness is grounded at both ends. Other than this it is OK. ICU4 CNP2 HSL Pin 7 missing. Second screen is not fitted. This is as per FM cable. All EMC harnesses are acceptable. 15:00 Icu, structure, egse connected using emc cables. 17:45 start FFT no2 completed 18:15 start roe ft completed 19:04 start mhc ft completed 20:19 disable watchdog timer before running 40s qcm test (bytes 0x76, 0xc0, 0xa9, 0, 2, 0xff, 0xff) sequence aborts before turning off qcms so turn off maually: (bytes 0x70, 0, 0x96, 0, 4, 0, 0, 0, 1) enable watchdog: (bytes 0x76, 0xc0, 0xa9, 0, 2, 0, 1) 22:07 standby icu off egse off ************************************************* Wed 25 May 2005 JAT, MCRW 08:30 Egse booted. FSRUN0193 08:40 icu on eeprom 0 standby 0.70A manual 0.86A exit default run seq.24, ll.34 sequence checksum error remake seq.24 reload seq.24 run seq24 md_data.0001 rerun md_data.0002 md_data.0002 - md_data.0001 gives blank image. Good! 49 ss_pos 0xC040 48 cm_pos 0xAAC0 1 pzt_sgo 0x0078 55 pzt_drv 0x3E57 52 gra_s_pos 0 08:54 Power down icu, egse change hsl harness (right hand jackpost (as you look at it) on cnp connector not engaging) 09:27 boot msslfp FSRUN0194 09:32 icu on eeprom 0 standby 0.70A manual 0.85A exit default run seq24 md_data.0001 rerun md_data.0002 md_data.0002 - md_data.0001 gives blank image. Good! md_data.0001 - fsrun0193/md_data.0002 gives blank image. 49 ss_pos 0xC040 48 cm_pos 0xAAC0 1 pzt_sgo 0x0078 55 pzt_drv 0x3E41 52 gra_s_pos 0 09:50 run eeprom_dump4_7 mem_data.0000 10:06 standby icu off new run FSRUN0195 10:09 start FFT 25_may_05_no1 10:43 sft completed 11:29 roe ft finished 12:30 mhc ft finished 14:13 fft finished icu off mdp halt 15:34 new run fsrun0196 start SFT 25 may 05 no2 16:37 start FFT 25 may 05 no2 17:12 start ROE ft 17:57 start mhc ft 19:35 mhc ft finished 21:16 fft finished icu off 21:19 egse off ********************************* Thuraday 26 May 2005 JAT, MCRW 08:45 Boot egse computers FSRUN0198 09:05 icu on boot 0.29A eeprom 4 stdby 0.70A manual exit default 0.85A copy mhc code from eeprom 7 run mhc code init mhc mhc htrs turned off move grating reverse 100 steps nack (ec 0400) because runtime and steps not set re-enable trace buffer grating vane test - ok finished 09:52 10:22 mhc soft reset mhc init14 run mhc_htr_off 10:30 Coarse mirror auto movement testing (from 0xAADD) bytes 0x6c,0x28,0x27,0,2,0xb2,0x51 bytes 0x6c,0x28,0x27,0,2,0xa3,0x75 bytes 0x6c,0x28,0x27,0,2,0xAA,0xDD run trace_dump (11 - 21) 10:35 new_st st_data.0003 slit slot auto testing bytes 0x73,0x60,0x9f,0,4,0,1,0,0 bytes 0x73,0x60,0x9f,0,4,0,1,0,1 run trace_dump (22 - 32) mhc_reset abort 10:53 new_st st_data.0004 run eeprom_dump0_3 mem_data.0033 memxtract mem_data.0033 | mem_recon1.pl 0xc8000 > mem33.recon diff mem33.recon /data/eis/fm/icu/memory_maps/fm20/eeprom/eeprom0/fm20_pad0_new.ecmds This shows differences: mem33.recon contains eeprom 3 data while fm20_pad0_new.ecmds doesnt. This was because fm20_pad0_new.ecmds writes to eerpom 0-2, and eeprom 3 was written later with eeprom_3 commands. So to replace eeprom 0 -3 do: run fm20_pad0_new_ecmds run eeprom_3 11:52 new_st st_data.0005 load seq.01, ll.01 sel seq 1 bytes 0x22 (disable mode change) eis_mode_auto failed tc packet (id 0x21, error code 4 [invl mode]) mode enable auto shutter abort (no init mhc after mhc reset/abort) allowed mode change manual status clear edit ll.01 to make y_w 511 make and load run fm_integration/mhc_init14 sel seq 1 auto ll xsum abort manual revert ll.01 make and load auto md_data.0001 manual change x_len window 3 from 1024 to 1020 make ll01 load and go ll xsum abort stat1 0x22 & 0x23 AAA2 revert ll.01 make and reload auto actually loaded seq01 manual load ll01 auto md_data.0002 ok 12:11 stdby icu off power supply off mdp halt new run started FSRUN0199 lunch ------------------- 13:23 icu on boot 0.29A eeprom 4 stdby 0.70A relocate eeprom 4-7 to 0-3 run eeprom_dump0_3 mem_data.0000 13:47 new run FSRUN0200 Start FFT 14:20 finish sft 15:00 finish roe ft 15:57 finish mhc ft 17:35 finish fft icu off new run FSRUN0201 18:02 icu on boot standby (eeprom 0) manual exit default bytes 0x2c,3 load mhc from eeprom 3 run from ram/abort init mhc run roe_prescan_test md_data.0001,2,3 run roe_prescan_test md_data.0004,5,6 new_st st_data.0001 18:26 set_delay 500.0 run fm20_eeprom0_3 19:02 run eeprom_dump0_3 mem_data.0000 19:17 memxtract mem_data.0000 | mem_recon1.pl 0xc8000 > mem0.recon [solarb@msslqa fsrun0201]$ diff mem0.recon ../fsrun0198/mem33.recon [solarb@msslqa fsrun0201]$ No differences with previous fm20 contents in eeprom 0-3 19:22 standby icu off new run FSRUN0202 run sft using eeprom 0 code 19:43 sft finished ok new_st st_data.0001 select eeprom 4 standby manual, exit default bytes 0x2c,7 to load mhc code run from ram run mhc_init14 run 'fm_integration/roe_prescan_test' fsrun0202/md_data.0002,3,4 load seq24, ll34 auto fsrun0202/md_data.0005 rerun fsrun0202/md_data.0006 6-5=blank 20:02 standby icu off egse off Data backup Pack tomorrow ********************************************** Thursday 11 August 2005 HH, JAT, MCRW EIS Integration 2 testing. 10:05 Problems with connecting using SDTP in real time. Using J-side system to check housekeeping. Keichii saved the day (tlmuser_dp mode set to 2 (= late buffering)). Copy tlmuser_dp.c tlmuser_real.c and re-run. Recieve hk ok in real time. Updating problems still? Cal leds set to the old software value ... 12:06 Test finished ***********************************************